• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO, NORDICS; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO, NORDICS; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number ND2110X15B
Medical Device Problem Codes Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy evo, nordics ventilator was returned to a third-party service center for preventative maintenance.There was no harm or injury reported.The device was not in patient use.During the evaluation of the trilogy evo, nordics device at third-party service center, an evt_internal_batt_not_detected error was observed in the device's event log.The device's system board was replaced to address the issue.Additionally, the blower assembly, bellow blower, machine flow sensor, and in-line filter prox.Are contaminated and needed to be replaced.The air foam inlet filter, muffler assembly, and particulate filter were replaced due to preventive maintenance.This device has been serviced, inspected, tested, and met acceptance criteria in accordance with all of the applicable customer requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO, NORDICS
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23479645
Report Number2518422-2025-111607
Device Sequence Number10252318
Product Code CBK
UDI-Device Identifier00606959055513
UDI-Public(01)00606959055513
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberND2110X15B
Device Catalogue NumberND2110X15B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2024
Initial Report FDA Received Date11/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/25/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-