• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INFINITY TIBIAL TRAY SZ 4 LNG TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INFINITY TIBIAL TRAY SZ 4 LNG TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 33650014
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Arthritis (1723); Cyst(s) (1800); Post Operative Wound Infection (2446); Inadequate Osseointegration (2646)
Date of Event 10/07/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information has been requested and, if received, will be provided in the supplemental report.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery.The primary implant consisted of a tibial tray and talar dome.The patient later required removal of the implant and placement of an antibiotic cement spacer due to suspected infection, along with noted poor bone quality and a large bone cyst superior to the tibial tray.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery.The primary implant consisted of a tibial tray and talar dome.The patient later required removal of the implant and placement of an antibiotic cement spacer due to suspected infection, along with noted poor bone quality and a large bone cyst superior to the tibial tray.
 
Additional Manufacturer Narrative
The complaint was confirmed, since the information for evaluation matches the alleged failure.Upon further investigation of the ct scans by healthcare professionals the following was observed: the ct scan shows a girdlestone situation with a cement spacer at the former ankle joint.The initial operation was performed in 2016, infection was suspected, although not confirmed by biological findings.There is the large cysts visible adjacent to the cement which have been reported and the poor bone status can be confirmed.The date of implantation is approx.9 years ago and therefore a direct link to either the device or the implantation cannot be made.The failure is therefore not to be assessed as being linked to this.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY TIBIAL TRAY SZ 4 LNG TOTAL ANKLE SYSTEM
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23480143
Report Number3010667733-2025-01281
Device Sequence Number10252341
Product Code HSN
UDI-Device Identifier00840420122551
UDI-Public00840420122551
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140749
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number33650014
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/07/2025
Supplement Date Received by Manufacturer12/01/2025
Initial Report FDA Received Date11/05/2025
Supplement Report FDA Received Date12/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-