• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 4.5PCF
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that there was a damage to the tracheostomy tube's inflation line near the incision hole site during use.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3 and h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the inflation line was broken at the flange connection.The inflation line was partially broken at the connector flange joint, and part of the inflation line tubing remained bonded into the flange.Functionally, an inflation/deflation test was performed using a syringe.Air was applied to the cuff and it was observed that the cuff deflated immediately.It was reported that there was a damage to the tracheostomy tube's inflation line near the incision hole site during use.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.The instructions included with this device provide the following guidance: do not use larger than a 1.0 cc/ml syringe when adding air into the cuff and monitor the cuff pressures frequently.Each tube¿s cuff, pilot balloon and valve should be tested by inflation prior to tube insertion.If dysfunction is detected in any part of the inflation system, the tube should not be used.Do not overinflate the cuff.Ordinarily, cuff pressure should not exceed 25 cm of h2o.Closely monitor cuff pressure under a physician¿s guidance.Overinflation may lead to permanent tracheal damage, rupture of the cuff with subsequent deflation, or cuff distortion that may cause airway blockage.Avoid pulling or manipulation of the cuff inflation line as this could cause cuff deflation or removal of the tube from patient.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23480218
Report Number9681384-2025-01173
Device Sequence Number18701517
Product Code JOH
UDI-Device Identifier20884521779713
UDI-Public(01)20884521779713(17)300122(10)25A0872JZX
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K182861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4.5PCF
Device Catalogue Number4.5PCF
Device Lot Number25A0872JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2025
Supplement Date Received by Manufacturer02/19/2026
Initial Report FDA Received Date11/05/2025
Supplement Report FDA Received Date03/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-