SYNTHES GMBH RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
| Catalog Number |
05.001.240 |
| Medical Device Problem Codes |
Physical Resistance/Sticking (4012); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: section 10: general condition, section 20: markings, section 30: leakage test using bubble emission technique, section 80: check for sticky trigger, section 140: check general function of device.During the service evaluation the following failures were identified: visual : cosmetic damage - worn , functional : will not run ¿ trigger defective , internal finding : leak tightness test failure , labeling : illegible etch, functional : sticky trigger.Conclusion: failure reported is confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the recon sagittal saw with key for trs device was had a sticky trigger it was noted in the service order that the device did not work.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: upon further review of the device investigation, it was determined that the reported malfunction of the device having a sticky trigger was incorrect.It was determined that the malfunction did not occur and therefore, is unlikely to cause or contribute to serious injury or death if it were to recur.Therefore, this event is not reportable.
|
| |
|
Search Alerts/Recalls
|
|
|