| Model Number |
9799 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Masimo has reached out to the customer to request the return of the device.The device has not been returned to masimo to allow an analysis to be performed.If the device is returned for evaluation or new information is obtained, a follow up report will be submitted.Health effect impact code: no adverse event; no patient impact h3 other text : product not returned.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device keeps shutting off.There was no patient impact or consequence reported.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device keeps shutting off.There was no patient impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: the returned device was evaluated.The device was found to have a misaligned ib/conb flex cable.This prevents the device from completing the power-on sequence followed by an alarm state of 1 long beep and 3 short beeps.
|
| |
|