ST. JUDE MEDICAL, INC.(CRM-SYLMAR) OPTISURE ACTIVE FIXATION; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number LDA210/65
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 10/31/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had infection and endocarditis.The physician explanted the active system ¿ implantable cardioverter defibrillator, right ventricular lead, left ventricular lead and right atrial lead due to the infection.The previously capped right ventricular lead was also explanted.The patient was in stable condition.
 
Additional Manufacturer Narrative
The sterilization records were reviewed and no evidence of abnormal sterilization cycle was found.The results of the investigation are inconclusive since the lead was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Additional Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.The cause of infection could not be traced to the device.
 
Additional Manufacturer Narrative
Correction: section g3 in the previous supplemental report should have been 16 dec 2025 and not 5 dec 2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTISURE ACTIVE FIXATION
Common Device Name
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key23482497
Report Number2017865-2025-1004315
Device Sequence Number18699085
Product Code NVY
UDI-Device Identifier05414734507172
UDI-Public(01)05414734507172(10)A000146435(17)260731
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberLDA210/65
Device Lot NumberA000146435
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2025
Supplement Date Received by Manufacturer12/05/2025
12/19/2025
Initial Report FDA Received Date11/05/2025
Supplement Report FDA Received Date12/16/2025
12/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
Patient Weight103 KG
-
-