| Model Number |
LDA210/65 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
10/31/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had infection and endocarditis.The physician explanted the active system ¿ implantable cardioverter defibrillator, right ventricular lead, left ventricular lead and right atrial lead due to the infection.The previously capped right ventricular lead was also explanted.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The sterilization records were reviewed and no evidence of abnormal sterilization cycle was found.The results of the investigation are inconclusive since the lead was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.The cause of infection could not be traced to the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: section g3 in the previous supplemental report should have been 16 dec 2025 and not 5 dec 2025.
|
| |
|