Catalog Number 1809661
Medical Device Problem Codes Fracture (1260); Material Separation (1562); Migration (4003)
Health Effect - Clinical Codes Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/24/2025
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2025, a patient underwent a chemotherapy by using powerport mri isp.Prior to the injection of the contrast, it was noted through the fluoroscopy, the catheter of the port was broken half and migrated into the left pulmonary artery and the broken part was retrieved by the radiologist.Further patient went through serious pain.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H11: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of sample is pending.The investigation of the reported event is currently underway.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2025, a patient underwent a chemotherapy by using powerport mri isp.Prior to the injection of the contrast, it was noted through the fluoroscopy, the catheter of the port was broken half and migrated into the left pulmonary artery and the broken part was retrieved by the radiologist.Further patient went through serious pain.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H11: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.The investigation is inconclusive for the reported fracture, material separation and migrations as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.E1, g3, h6 (component, method, result, conclusion).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT MRI ISP W/O SUTURE PLUGS, 9.6 FR SILICONE SL CATHETER
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key23483501
Report Number3006260740-2025-08440
Device Sequence Number9510232
Product Code LJT
UDI-Device Identifier00801741027161
UDI-Public(01)00801741027161
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K073423
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 12/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1809661
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2025
Supplement Date Received by Manufacturer12/24/2025
Initial Report FDA Received Date11/06/2025
Supplement Report FDA Received Date12/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexMale
Patient Weight89 KG
Patient RaceWhite
-
-