• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 52E; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 52E; PROSTHESIS, HIP Back to Search Results
Catalog Number 010000663
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Metal Related Pathology (4530)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: unknown 36mm liner.Unknown head.Unknown taperloc microplasty size 6 stem.Unknown sleeve.The customer indicated that the product remains implanted; therefore, it will not be returned to zimmer biomet for investigation.Once the investigation has been completed, a supplemental mdr will be submitted.
 
Event or Problem Description
It was reported that the patient has been experiencing right hip issues and pain since his initial implantation.It is now approximately 11 (eleven) years post-surgery.No further details or medical records have been provided.
 
Event or Problem Description
No additional information available for the reported event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b6; d1; d2b; d4; d10; g3; g4; h2; h4; h6.D10: g7 neutral e1 liner 36mm e.Item: 010000857.Lot: 3302871.Tprlc 133 mp type1 pps so 11.0.Item: 51-106110.Lot: 3311802.Cer bioloxd option hd 36mm.Item: 650-1057.Lot: 392850.Cer option type 1 tpr sleve -6.Item: 650-1064.Lot: 992340.Once the investigation has been completed, a supplemental mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: a5; a6; b5; b6; b7; g3; h2; h3; h6; h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: patient had a right tha previously.Patient has since been having issues and reporting pain in the hip.During the follow-ups the doctor notes he is concerned for metallosis and trunnionosis.Patient had x-rays and blood lab results.Results have no evidence of fracture or dislocation, had good alignment.Lab test results for chromium and cobalt are normal within range.Mri results show no evidence of osseous abnormality or osteolysis.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor trends.
 
Event or Problem Description
It was reported that the patient has been experiencing right hip issues and pain since his initial implantation.It is now approximately 11 (eleven) years post-surgery.An mri was performed that displayed small chronic bone fragments at the greater trochanter; otherwise, unremarkable without evidence of osseous abnormality or osteolysis.All products remain implanted.Due diligence is complete as multiple attempts were made; however, no further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACET SHELL 52E
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23485636
Report Number0001825034-2025-03490
Device Sequence Number10559129
Product Code PBI
UDI-Device Identifier00880304524224
UDI-Public(01)00880304524224(10)3360499(17)240725
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/25/2024
Device Catalogue Number010000663
Device Lot Number3360499
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2025
Supplement Date Received by Manufacturer11/11/2025
01/20/2026
Initial Report FDA Received Date11/06/2025
Supplement Report FDA Received Date12/03/2025
01/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/25/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10; SEE H11
Outcome Attributed to Adverse Event Other;
Patient Age64 YR
Patient SexMale
Patient Weight88 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-