• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STORZ MEDICAL AG MODULITH SLX-F2; 25000

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STORZ MEDICAL AG MODULITH SLX-F2; 25000 Back to Search Results
Model Number 25000
Medical Device Problem Codes Misfocusing (1401); Use of Device Problem (1670)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 10/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Hematoma are known side effect of eswl.Focal point check (7.2 in maintenance record) did not pass.For targeted shockwave application, proper alignment between lithotripter and x-ray system is required.The user is thus advised (training/ifu) to perform a focus check daily before treatment.The ifu also states that "no treatment is allowed to be performed if the reflection mark is not within the specified area.Inform your service centre." we did ask the user if they had followed our instructions and they said they would usually perform the focus check tri-monthly as part of their qa.According to the user, the last qa was performed on 9.9.2025, i.E., after the occurrence of the first reported hematoma and before the remaining two hematomas.The hospital/user of the device has been reminded to perform all relevant function checks as per our instructions.Misalignment has been corrected and device was put back into service.
 
Event or Problem Description
On (b)(6) 2025 ctu post eswl treatment confirmed large subcapsular hematoma.Admitted on (b)(6) 2025 for observation, no transfusion or other intervention, discharged on (b)(6) 2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULITH SLX-F2
Common Device Name
25000
Manufacturer (Section D)
STORZ MEDICAL AG
lohstampfestrasse 8
taegerwilen, 8274 che
SZ 
Manufacturer (Section G)
STORZ MEDICAL AG
lohstampfestrasse 8
taegerwilen, 8274 che
SZ  
Manufacturer Contact
devonne pavlak
1000 cobb place blvd.
building 400 suite 450
kennesaw, GA 30144
6785817255
MDR Report Key23486490
Report Number9613347-2025-00009
Device Sequence Number18724513
Product Code LNS
UDI-Device Identifier07630039100230
UDI-Public07630039100230
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K072788
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2025
Initial Report FDA Received Date11/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age43 YR
Patient SexFemale
-
-