• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD CARIBE LTD. BD MICROTAINER® PST¿ TUBES WITH LH (LITHIUM HEPARIN); BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD CARIBE LTD. BD MICROTAINER® PST¿ TUBES WITH LH (LITHIUM HEPARIN); BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 365985
Medical Device Problem Code Improper Chemical Reaction (2952)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
Report 3 of 4: it was reported during use of bd microtainer® pst¿ tubes with lh (lithium heparin), 1 patient sample was hemolyzed.There was no health impact or consequences reported.
 
Event or Problem Description
Report 3 of 4: it was reported during use of bd microtainer® pst¿ tubes with lh (lithium heparin), 1 patient sample was hemolyzed.There was no health impact or consequences reported.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for evaluation: yes.H3: device eval by manufacturer? yes.D9: returned to manufacturer on: 17-oct-2025.Investigation summary: bd received two hundred and fifty unused representative customer samples for investigation.The samples were visually evaluated.Although the reporting condition of hemolysis was not confirmed, gel smear was observed on eight samples.The barrier gel was smeared on the wall of tube and located towards the top of the tube.Additionally, fifty (50) retention samples were visually inspected for additive presence and gel defects and no issues were observed.Complaints for sample quality are under statistical control for the month of october 2025.At this time, further testing is not indicated based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has not been confirmed for hemolysis but has been confirmed for gel smearing based on the customer samples provided.Bd was not able to identify a root cause for gel smearing.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® PST¿ TUBES WITH LH (LITHIUM HEPARIN)
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos PR
Manufacturer (Section G)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos PR
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
2018475186
MDR Report Key23487576
Report Number2618282-2025-00131
Device Sequence Number10488014
Product Code JKA
UDI-Device Identifier30382903659853
UDI-Public(01)30382903659853(17)260131(10)431641N
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K991702
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2026
Device Catalogue Number365985
Device Lot Number431641N
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2025
Supplement Date Received by Manufacturer11/14/2025
Initial Report FDA Received Date11/06/2025
Supplement Report FDA Received Date11/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-