The manufacturer received information alleging a broken fan.There was no serious patient harm or injury that occurred or was reported.The device was not in patient use.During the evaluation of the device at third-party service center, a ventilator inoperative error code related to the machine flow sensor was found in the ventilator's downloaded error log.Evt_mach_flow_drift_high means the machine flow sensor drifted above the specification (>0.2lpm).The flow sensor was replaced to address the issue.Additionally, un-related to the reportable malfunction/error code, an error code related to the motor controller and an error code related to sensor offset were found in the ventilator's downloaded error log.Evt_mcl_foc_shutdown means the motor controller has proceeded to the shutdown state due to an error with the motor.The blower was replaced to address the issue.Evt_drift_debug means the sensor offset during drift calculation is too high.The circuit board was replaced to address the issue.Also, one of the in-line filters prox.Was replaced due to contamination.
|