• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number EE2110X15B
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a broken fan.There was no serious patient harm or injury that occurred or was reported.The device was not in patient use.During the evaluation of the device at third-party service center, a ventilator inoperative error code related to the machine flow sensor was found in the ventilator's downloaded error log.Evt_mach_flow_drift_high means the machine flow sensor drifted above the specification (>0.2lpm).The flow sensor was replaced to address the issue.Additionally, un-related to the reportable malfunction/error code, an error code related to the motor controller and an error code related to sensor offset were found in the ventilator's downloaded error log.Evt_mcl_foc_shutdown means the motor controller has proceeded to the shutdown state due to an error with the motor.The blower was replaced to address the issue.Evt_drift_debug means the sensor offset during drift calculation is too high.The circuit board was replaced to address the issue.Also, one of the in-line filters prox.Was replaced due to contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23487598
Report Number2518422-2025-111635
Device Sequence Number10488442
Product Code CBK
UDI-Device Identifier00606959055544
UDI-Public(01)00606959055544
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEE2110X15B
Device Catalogue NumberEE2110X15B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2025
Initial Report FDA Received Date11/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/23/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-