• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number DDU-A2300
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that their aed's asi is flashing red, and reporting a service code.The customer did not report this occurring during patient use.
 
Additional Manufacturer Narrative
Although the customer initially reported a service code, review of the aed's internal log files determined that the service code after battery replacement was due to the previous battery fully depleting within the unit.Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.As a follow up with the customer, the proper maintenance of this device was reviewed.Once the new battery was installed and a manual self test performed the aed functioned as designed and was able stay in service.This mdr is being submitted due to the aed becoming non-operational as a result of the battery reaching complete depletion without evidence of user intervention or maintenance following the aeds red asi indicating the unit needs attention or servicing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford CT 06405
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key23487799
Report Number3003521780-2025-01930
Device Sequence Number16074329
Product Code MKJ
UDI-Device Identifier10815098020208
UDI-Public(01)10815098020208
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-A2300
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2025
Initial Report FDA Received Date11/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-