| Catalog Number |
V072228 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the wound drains are too compressible and stop working.They have broken off in patients, and they have caused seromas in patients.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states: to avoid the possibility of drain damage or breakage: - avoid suturing through drains.- drains should lie flat and in line with the skin exit areas.- particular care should be taken to avoid any obstacles to the drain exit path.- drains should be checked for free motion during closure to minimize the possibility of breakage.- drain removal should be done gently by hand.Drains should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.- surgical removal may be necessary if drain is difficult to remove or breaks.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the wound drains are too compressible and stop working.They have broken off in patients, and they have caused seromas in patients.
|
| |
|