• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. 145-DEG PE 38MM HUM LINER +0; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. 145-DEG PE 38MM HUM LINER +0; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-38-00
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Radiolucent Lines (4848)
Date of Event 10/21/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial tsa (total shoulder arthroplasty) on the right side.Subsequently, the patient experienced pain.As a result, the patient was revised for lucency of the stem and infection.Once parts were removed during the case, it appeared the lip of the poly had been flattened down and eroded away.Reverse shoulder components were removed, and an antibiotic spacer was re implanted.Patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
D10: (b)(6), 300-01-12 - equinoxe humeral stem primary press fit 12mm.(b)(6), 320-01-38 - equinoxe reverse 38mm glenosphere.(b)(6), 320-10-00 - equinoxe reverse tray adapter plate tray +0.(b)(6), 320-15-01 - eq rev glenoid plate.(b)(6), 320-15-05 - eq rev locking screw.(b)(6), 320-20-00 - eq reverse torque defining screw kit.(b)(6), 320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm.(b)(6), 320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm.(b)(6), 320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm.(b)(6), 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm.(b)(6), 321-20-00 - equinoxe reverse shoulder drill kit.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/technique.A definitive root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
145-DEG PE 38MM HUM LINER +0
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66 ct
gainesville, FL 32653
3523771140
MDR Report Key23488267
Report Number1038671-2025-03282
Device Sequence Number13325096
Product Code KWT
UDI-Device Identifier10885862086655
UDI-Public10885862086655
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/11/2024
Device Catalogue Number320-38-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2025
Supplement Date Received by Manufacturer02/12/2026
Initial Report FDA Received Date11/06/2025
Supplement Report FDA Received Date02/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/14/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization;
Patient Age81 YR
Patient SexFemale
-
-