|
The mayfield swivel adaptor (a1018) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the inspection of the returned swivel adapter shows that one retainer ring and washer was missing.To resolve all issues, the swivel adapter¿s nylon washers and the retaining rings were replaced, and the unit was reassembled.The functional test was carried out successfully and the device meets manufacturer specifications.Root cause - the complaint is confirmed via inspection of the unit.The swivel adapter was missing a retaining ring and washer.The probable root cause is mishandling or improper disassembly.No further corrective action is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|