| Model Number |
TB-0535FCS |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the subject device tissue pad partially detached.The issue occurred during a therapeutic left laparoscopic nephrectomy procedure.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection, and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: "teflon detachment" was confirmed and a defect in the forceps is determined.The most probable cause of the discovered defect is d02, cause traced to component failure, which is expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|