This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h3, h6 the device was returned to olympus for inspection and the reported failure was confirmed.The probe was broken off at approximately 17 mm from its distal end.A root cause could not be identified.Based on the results of the investigation, it is likely the following mechanism led to the malfunction: 1.During the seal and cut output, the probe came into contact with hard tissue, metal, or surgical equipment, causing a scratch.2.The probe was subjected to the load of the seal and cut output or the gripping of tissue, causing a crack to form from the scratch.3.The probe broke due to the application of load.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|