| Model Number |
M00545180 |
| Medical Device Problem Codes |
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices that were used in the same patient and procedure.It was reported to boston scientific corporation that an autotome rx 44 was attempted to be used in the papilla of vater during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat common bile duct (cbd) stones performed on (b)(6) 2025.During the procedure, the dreamtome rx 44 weas used but, the cutting wire broke the moment the wire was energized.An autotome rx 44 was attempted to be used but, the cutting wire broke.It was reported that no part of the cutting wire of either device detached or fell into the patient.The procedure was completed with another autotome rx 44.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire break.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire break.Block h11: (investigation result).The returned autotome rx 44 was analyzed, and a visual evaluation noted that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Media evaluation noted that the device inside the pouch with the cutting wire broken and kinked.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Based on the condition of the device, the wire break could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices that were used in the same patient and procedure.It was reported to boston scientific corporation that an autotome rx 44 was attempted to be used in the papilla of vater during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat common bile duct (cbd) stones performed on (b)(6) 2025.During the procedure, the dreamtome rx 44 weas used but, the cutting wire broke the moment the wire was energized.An autotome rx 44 was attempted to be used but, the cutting wire broke.It was reported that no part of the cutting wire of either device detached or fell into the patient.The procedure was completed with another autotome rx 44.There were no patient complications reported as a result of this event.
|
| |
|