Model Number FX834T
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/20/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The adverse event is filed under aic reference (b)(4).Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
According to the complainant the shunt was explanted because cerebrospinal fluid (csf) was accumulating.Additional information has been requested but not yet provided.
 
Additional Manufacturer Narrative
The adverse event is filed under aic reference (b)(4).Additional information: a2 patient age.A3a gender.B7 relevant history.D item information updated.D6a, d6b implant and explant date.D9 device returned to manufacturer date.Visual inspection: during the investigation, no significant deformations or damages of the valves were determined.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The result shows that the m.Blue operates not within the accepted tolerances in the horizontal position.However, the progav 2.0 operates within the specified tolerances.An slower outflow of m.Blue could be determined.Adjustment test: the valves were tested and both valves are adjustable in all pressure settings.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerance.Internal inspection: after dismantling of the valves, organic deposits were found in both valves.Results: according to the results of the investigation, a reduced outflow of the m.Blue in the horizontal position was detected.The visible deposits led to the functional deviation.Furthermore, visible deposits in the progav 2.0 were determined.The deposits had no effect upon the specifications at the time of the examination.The valve operated within all specifications.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.The examination of the return has been completed with the drafting of this report.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE PLUS SYS W/SPRUNG RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT,
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469 ,
GM  14469,
MDR Report Key23496104
Report Number2916714-2025-00072
Device Sequence Number18420835
Product Code JXG
UDI-Device Identifier04041906504854
UDI-Public(01)04041906504854(11)240531(17)290521(10)20075598
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/23/2025
Report Date (Section F) 11/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX834T
Device Catalogue NumberFX834T
Device Lot Number20075598
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/20/2025
Device Age19 MO
Event Location Hospital
Date Report to Manufacturer10/20/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/07/2025
Supplement Report FDA Received Date12/23/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age1 MO
Patient SexMale
Date Report Sent to FDA11/07/2025
-
-