| Model Number |
BURR ATTACHMENT 2.35 MM |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 10/21/2025, a facility representative reported via (b)(4) that a burr broke off into an ar-300b attachment, and the stem was stuck inside, preventing another attachment from connecting.Device swapped and case completed with no patient effect.Additional info provided 10/31: ar-300-b102 burr was broken during procedure, no adverse effects or fragments produced from breakage.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on photographic evidence of a damaged ar-300b batch 14955784 the omplaint allegation is confirmed.Based on the information provided ¿ which may include the returned device (if available), photographs, videos, event description, and any additional field input ¿ arthrex has determined the most likely cause.The most likely cause is attributed to user-related factors such as mechanical overstress at the burr¿attachment interface.
|
| |
|