| Model Number |
989605449841 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/09/2025
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that the lumify ultrasound system was not available during an emergent ecmo (extra-corporeal membrane oxygenation) because the transducer was not registered to the system, which resulted in a patient death.Additional information is being obtained.Philips has started an investigation, and upon completion, a follow up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed, including manual reproduction testing using another device to identify the cause of the reported issue.The engineering team determined that the transducer usb-c connector can exhibit reduced mechanical retention when mated to the tablet port.This can lead to an intermittent issue where the firmware version is unrecognized.When this occurs a prompt to reregister the transducer is displayed.
|
| |
|