| Model Number |
DDPB3D1 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
10/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 1688t lead implanted: (b)(6) 2024; 1688t-52 lead implanted: (b)(6) 2007.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient received possible inappropriate high voltage therapy and anti-tachycardia pacing (atp) for supra ventricular tachycardia (svt) versus ventricular fibrillation (vf). it was noted that therapies were withheld by atrial fibrillation (af)/aflutter discriminator initially and at time of therapy at/af event was in progress. the implantable cardioverter defibrillator (icd) remains in use. it was also reported that the right ventricular (rv) lead exhibited high thresholds. the rv lead remains in use. no further patient complications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
It was further reported that the crt-d high voltage therapy and atp was appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|