Catalog Number 175814
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the foley catheter was placed for scheduled cesarean section.Attempted to inflate foley balloon, fluid from syringe went directly to foley bag, thus the balloon could not be inflated.
 
Event or Problem Description
It was reported that the foley catheter was placed for scheduled cesarean section.Attempted to inflate foley balloon, fluid from syringe went directly to foley bag, thus the balloon could not be inflated.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states".Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Proper techniques for urinary catheter maintenance: maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.Maintain unobstructed urine flow and keep the catheter and collection tube free from kinking keep the collection bag below the level of the bladder or hips at all times.Empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient.Routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate.Leave foley catheter in place only as long as needed.Directions for use.13.Proceed with catheterization in usual manner using the dominant hand a.When catheter tip has entered bladder, urine will be visible in the drainage tube b.Insert catheter two more inches and inflate catheter balloon.14.Inflate catheter balloon using entire 10cc of sterile water provided in the prefilled syringe.Note: use of less than 10cc can result in asymmetrically inflated balloon.15.Once inflated, gently pull catheter until the inflated balloon is snug against the bladder neck.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Correction: d.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER
Common Device Name
SILICONE FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23497745
Report Number1018233-2025-10072
Device Sequence Number8993371
Product Code EZL
UDI-Device Identifier00801741034152
UDI-Public(01)00801741034152(11)250403(17)270731(10)NGKP4405
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K984084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number175814
Device Lot NumberNGKP4405
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/31/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2025
Supplement Date Received by Manufacturer12/18/2025
06/17/2026
Initial Report FDA Received Date11/07/2025
Supplement Report FDA Received Date01/07/2026
06/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-