MAUDE Adverse Event Report: MICROVENTION, INC. LVIS D; INTRACRANIAL COIL-ASSIST STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. LVIS D; INTRACRANIAL COIL-ASSIST STENT Back to Search Results
Model Number 213041-LVIS-D-PMA
Medical Device Problem Codes Material Twisted/Bent (2981); Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2017
Type of Reportable Event Serious Injury
Event or Problem Description
As reported via the lvis pas clinical study: this file captures the second stent used in the case.During index procedure on (b)(6) 2017, microwire was then removed and the first of the two flow diverting stents was then loaded and applied in the following manner.First part of the stent was pushed out of the catheter such that good distal wall apposition was obtained.The rest of the stent was then gently pushed out of the catheter buckfield open completely in the supra-clinoid ica segment, then "wagged" the construct a few times until it opened fully and continued deploying the remainder of the device.Follow up runs reveal incomplete stent apposition along the anterior ophthalmic and clinoid segments.We then used a j-wire to bump the stent open and also gain distal access for the second device.Follow up angiography after this maneuver revealed complete apposition of the stent but persistent aneurysm filling.We then loaded the second device and started deploying it but, yet again, had trouble with incomplete opening.The above-mentioned procedure ¿wagging¿ was repeated with only partial success.The device was continued being deployed but had to reach sheath it numerous times in the process of deployment and eventually were able to be landed with good distal and partial proximal wall apposition.It was decided to angioplasty through the stent especially in its proximal one third.A total of 4 balloon deployments were performed after which follow up angiography revealed satisfactory release of the stent and good wall apposition.There is one single tiny none of the stent which is looped back on itself but is opposed to the wall of the vessel instead of being luminal.Follow up angiography revealed good antegrade flow, no stent thrombosis, persistent but delayed filling of the aneurysm, no vascular dissections or vasospasm.
 
Additional Manufacturer Narrative
Based on a review of the data and in my professional opinion, the second lvis stent encountered incomplete opening and the relationship of this event to the lvis device cannot be ruled out.The physical device was not available for evaluation to investigate if a condition existed that would have contributed to the reported event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation further examine the cause of the reported event.This information may be updated if additional information is provided at a later date.Based on a review of the device¿s risk documentation, the reported event did not indicate there were the presence of any potential or new manufacturing, design, quality, or other systemic issues, or non-conformance's.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform an analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformance's associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential adverse events: the following potential risks and complications associated with general anesthesia, cerebral angiography, intracranial catheterization, intracranial stent placement or intra-saccular coil deployment have been identified below: allergic reaction, including but not limited to: contrast dye, nitinol metal, and any other medications used during the procedure, aphasia, blindness, cardiac arrhythmia, coil prolapsed or migration into normal vessel adjacent to aneurysm, complications of arterial puncture including pain, local bleeding, local infection and injury to the artery, vein or adjacent nerves, cranial neuropathy, death, device fracture, migration or misplacement, dissection or perforation of the parent artery, headache, hemorrhage (i.E., intracerebral hemorrhage (ich), subarchnoid hemorrhage (sah), or retroperitoneal (or in other locations), hemiplegia, hydrocephalus, infection, injury to normal vessel or tissue, ischemia, mass effect, myocardial infarction, neurological deficits, occlusion of non-target side branches, pseudo aneurysm formation, reactions to anti-platelet/anti-coagulant agents, reactions due to radiation exposure, reactions to anesthesia and related procedures, reactions to contrast agents, renal failure, aneurysm rupture, stenosis of stented segment, seizure, stent thrombosis, stroke or tia (transient ischemic attack), thromboembolic event (t/e), vasospasm, visual impairment.Warnings: should unusual resistance be felt at any time during access or removal, the introducer/guide catheter/microcatheter and lvis device should be removed as a single unit.Applying excessive force during delivery or retrieval of the lvis device can potentially result in loss or damage to the device and delivery components.It is imperative to use the lvis device with compatible microcatheters.If repeated friction is encountered during lvis device delivery, verify microcatheter is not kinked or in extremely tortuous anatomy.Confirm that the microcatheter does not ovalize.Confirm that there is adequate sterile flush solution.Do not reposition the lvis device in the parent vessel without fully retrieving the device.The lvis device must be retrieved into the microcatheter and re-deployed at the desired target location or removed completely from the patient.Precautions: exercise caution when crossing the deployed/detached lvis device with adjunctive devices such as guidewires, catheters, microcatheters or balloon catheters to avoid disrupting the device geometry and device placement.Directions for use 15.Advance the delivery wire to transfer the lvis device from within the introducer into the microcatheter.Warning: do not torque the delivery wire while advancing or retracting the lvis device.A torque device should not be used.16.Continue advancing the delivery wire into the microcatheter until the proximal tip of the delivery wire enters the introducer.Loosen the rhv locking ring, remove the introducer, and set it aside.Warning: do not apply undue force.If resistance is encountered at any point during lvis device delivery or manipulation, withdraw the unit and select a new lvis device.18.Position the lvis device for deployment by aligning the lvis implant distal radiopaque end markers approximately 7 mm past the aneurysm neck.[figure 6] note: a proper push/pull technique, encompassing sufficient delivery wire push force, in addition to an opposing microcatheter withdrawal force, will facilitate properly deploying the lvis device to achieve full expansion and good vessel apposition.Note: slowly advancing the lvis device while adjusting the microcatheter position will ensure accurate deployment.Maintain simultaneous control of the lvis device and microcatheter in order to position and expand the device at the proper location.Caution: using a rapid microcatheter withdrawal technique to deploy the lvis device is not recommended and may result in device elongation.19.If lvis device positioning is not satisfactory, the lvis device may be recaptured and repositioned if it is not fully deployed.The lvis device may be recaptured until the point where the proximal end of the lvis device markers is aligned 3 mm proximally with the microcatheter distal marker band (approximately 80% deployed).[figure 7] caution: if resistance is felt while recapturing the lvis device, do not continue to recapture the device.Withdraw the microcatheter slightly to unsheath the lvis device (without exceeding the recapture limit), and then attempt to recapture the lvis device.Caution: the lvis device must not be re-deployed more than three times.Note: the lvis device delivery wire should not be utilized as a guidewire after stent deployment.Do not torque the lvis device.A torque device should not be used.20.If lvis device positioning is satisfactory, carefully retract the microcatheter and advance the delivery wire together, to allow the lvis device to deploy across the neck of the aneurysm.Ensure the device proximal radiopaque end markers are approximately 7 mm proximal to the aneurysm neck to ensure an adequate landing zone.The lvis device will expand and total length may foreshorten up to 60% from its undeployed length (refer to tables 1) as it exits the microcatheter.Ensure microcatheter is retracted and clear from the proximal flared ends.Note: visualize and refer to the implant radiopaque end markers to maintain adequate implant length, approximately 7 mm on each side of the aneurysm neck or target location to ensure appropriate neck coverage.[figure 8] warning: do not detach the lvis device if it is not properly positioned in the parent vessel.Observe the delivery wire distal tip to assure it remains within the desired location of the parent vessel.21.Prior to removing the delivery wire and if necessary, carefully position the microcatheter distal to the lvis device to maintain access through the lvis device.Remove and discard the delivery wire.Warning: the lvis device delivery wire should not be utilized as a guidewire.Do not torque the lvis device.A torque device should not be used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LVIS D
Common Device Name
INTRACRANIAL COIL-ASSIST STENT
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key23499333
Report Number2032493-2025-90636
Device Sequence Number15661071
Product Code QCA
UDI-Device Identifier00842429100967
UDI-Public(01)00842429100967(11)161219(17)191130(10)16121925M
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P170013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2019
Device Model Number213041-LVIS-D-PMA
Device Lot Number16121925M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2025
Initial Report FDA Received Date11/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/26/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age1 MO
Patient SexFemale
-
-