Product complaint # (b)(4) h6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigation summary: the product was returned to ethicon for evaluation.One open box with one unopened overwrap packet was received for analysis.Product code 1667g.To evaluate the condition of the returned sample, the packet was opened.The paper cover was removed from the tray, and it was found to contain one single armed strand, which aligns with the requirements for product code 1667g.Additionally, the needle-suture combination corresponds to this product code.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no issues related to missing component / incorrect quantity were noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.
|
|
It was reported that a patient underwent an unknown surgery on an unknown date and suture was used.During the procedure, the product was supposed to be a single-armed suture, but it was a double-armed suture.It was not a control release needle.When we send the sample to you, we will let you know its return date and tracking number.Further details are not provided from the hp.No adverse patient consequences were reported.Additional information was requested.
|