| Catalog Number |
72200616S |
| Medical Device Problem Codes |
Overheating of Device (1437); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a procedure, it was noticed that the motor drive unit was overheating.The procedure was performed without any delay, using an equivalent s+n back up product.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of the device, and no physical damage was observed.A functional evaluation revealed a blade stall error and jammed motor.Further evaluation revealed the drive fork was stiff when rotated.It is noted the returned device grew warm during the functional evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with a component failure.Factors that could have contributed to the blade stall error and jammed motor include corrosion in the motor/gearbox assembly or a shorted phase of the motor.One of the causes of a shorted motor phase is a short in the power cord.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|
|
|