| Model Number |
VICM5_12.6(-6.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no additional similar complaint type events within associated lots were found.Refractive surprise - each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.Due to refractive surprise, the lens was exchanged for a same model and length lens.The problem was resolved.Cause of the event was reported as "patient related factor".There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.Additional information was requested.No further information is available at this time.This file will be re-opened if additional information is provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Originally the claim was determined to be reportable, but upon further review it was determined to be not mdr reportable, however an mdr was submitted.Claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|