| Model Number |
DDU-A2300 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon return, the device underwent bench testing.During testing it was determined that the aed was unable to successfully deliver a shock.Further investigation identified a connection issue within a circut board.Although the original customer report did not involve patient use and was not considered reportable, evaluation confirmed a malfunction that could delay or prevent therapy delivery in a clinical situation.As such, this event is being reported in accordance with 21 cfr 803.50.
|
| |
|
Event or Problem Description
|
|
A customer reported that their aed's asi is flashing red, and reporting a service code.The customer did not report this occurring during patient use.
|
| |
|