• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NG/NI FEEDING TUBE, STYLET WITH ENFIT CONNECTOR; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NG/NI FEEDING TUBE, STYLET WITH ENFIT CONNECTOR; DH CPK NG TUBES Back to Search Results
Model Number 42-7558
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Avanos medical received a single report that referenced five different incidences, which were associated with separate units, involving five different events.This is the fifth of five reports.Refer to 9611594-2025-00263 for the first event.Refer to 9611594-2025-00264 for the second event.Refer to 9611594-2025-00265 for the third event.Refer to 9611594-2025-00266 for the fourth event.It was reported the nasojejeunal (nj) tube was guidewire was noticed to be intermittently protruding when the tube was inside the patient.Unknown if injury occurred.Additional information received 22-oct-2025 stated "on flushing tube prior to insertion the guide wire protruded outside of the side feeding hole by 1-2mm.
 
Additional Manufacturer Narrative
The sample is reportedly available for this complaint but was not returned when this report was filed.A review of the device history record and udi number are in-progress.Upon completion of the sample evaluation and investigation; a follow-up report will be filed.All information reasonably known as of 06-nov-2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The actual complaint product was returned for evaluation.All information reasonably known as of 31-dec-2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30356064, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.The subject sample provided was evaluated confirming the stylet guidewire purple hub was seen engaged into the 2-port y-connector securely.At the same manner, the distal bolus end of the tube was inspected and did not observe the tip of the stylet guidewire went through the bolus plug window.When the stylet guidewire was re-inserted back into the yellow tubing, an appearance of stylet tip was visible through the bolus plug window.Root cause could not be determined.All information reasonably known as of 12-mar-2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NG/NI FEEDING TUBE, STYLET WITH ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key23503615
Report Number9611594-2025-00267
Device Sequence Number18418680
Product Code KNT
UDI-Device Identifier00350770460833
UDI-Public(01)00350770460833(240)104608302(17)280103(10)30356064
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number42-7558
Device Catalogue Number42-7558
Device Lot Number30356064
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2025
Supplement Date Received by Manufacturer11/11/2025
02/11/2026
Initial Report FDA Received Date11/07/2025
Supplement Report FDA Received Date12/31/2025
03/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-