|
An email was received regarding a 46 50 ml diluent set for channel 2, spiros, alleging a leak during patient use.It was stated in the report that there was leakage from the spiros.The event was noted during infusion.Moriamin-sn was involved in the event.There was no bleedback reported.The product is not a biohazard, and the device was not reprocessed/resterilized.There was patient involvement but no patient harm.There was no delay in critical therapy.
|
|
Received one (1) used.List #ch4002, 46" (117 cm) 50 ml, diluent set for channel 2 w/20 drop drip chamber, spiros¿; lot #13659713.No visual damages or anomalies were observed.The reported complaint of a leak could be confirmed.The returned sample was tested, and a leak was observed at the connection of the syringe and the transfer set.The sample was disconnected and found to have a broken luer in the syringe.The probable cause of the broken syringe is from an unintentional bending force during use.A device history review (dhr) of lot #13659713 and relevant commodities was reviewed, and the following discrepancy was found.Discrepancy: what? one unit, part number x1-4693 rev.01, lot #13753212, was reported to leak.When? september 13, 2023.Where? during the receiving inspection process, aql 0.025 c=0.Quantity impacted: 47,451 pieces.
|