• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INVISION TIBIAL TRAY SZ 3 +8MM UNIVERSAL LONG; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INVISION TIBIAL TRAY SZ 3 +8MM UNIVERSAL LONG; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 60003802
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Cyst(s) (1800); Fall (1848); Limb Fracture (4518)
Date of Event 10/13/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation is completed, any additional information will be reported in a supplemental report.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery of both the tibial and talar components because patient was walking, barefoot in a field, walking backwards and stepped in a pothole and twisted the ankle and broke the tibia.From then mal alignment began with the implant.
 
Additional Manufacturer Narrative
Correction - h6 (device code) the reported event could be confirmed, since the ct scan shows a periprosthetic posterior tibial fracture adjacent to the stemmed tibial component.It also looks as if some loosening was present for a longer time before the fracture occured as there is some radiolucence and cysts visible adjacent to the tibial component.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Upon further investigation of the ct scans, healthcare professionals opined that- the ct scan shows a periprosthetic posterior tibial fracture adjacent to the stemmed tibial component.It also looks as if some loosening was present for a longer time before the fracture occured as there is some radiolucence and cysts visible adjacent to the tibial component and the talar component looks somehow subsided.The underlying cause might be related to poor bone substance and therefore a patient related factor.However, failure of total ankle replacement tar over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information e.G., clinical status, exact symptoms, and range of motion of the patient, assessment of the treating physician is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and or the device in relation to cause of the failure.Based on investigation, the issue occurred most probably due to patient related factor i.E.Poor bone substance.More detailed information about the complaint event as well as the affected device must be available to determine the root cause of the complaint event.If the device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery of both the tibial and talar components because patient was walking, barefoot in a field, walking backwards and stepped in a pothole and twisted the ankle and broke the tibia.From then mal alignment began with the implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVISION TIBIAL TRAY SZ 3 +8MM UNIVERSAL LONG
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23507351
Report Number3010667733-2025-01297
Device Sequence Number15679574
Product Code HSN
UDI-Device Identifier00889797076548
UDI-Public(01)00889797076548(11)210731(17)290729(10)1693296
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K142117
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number60003802
Device Lot Number1693296
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/13/2025
Supplement Date Received by Manufacturer01/30/2026
Initial Report FDA Received Date11/09/2025
Supplement Report FDA Received Date02/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-