• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC. CORNERSTONE SR INSTRUMENTS; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC. CORNERSTONE SR INSTRUMENTS; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Model Number 875777
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: first name and last name of initial reporter is unknown g2: country of origin is japan h3: product analysis for part # 875777; lot # jm19g002 visual inspection confirmed the catalyst inserter handle has broken.The damage to the instrument is consistent with overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having cervical spine anterior fusion spinal therapy for cervical spondylotic myelopathy.Levels implanted at c4/5.It was reported that the handle had cracks during post-operative cleaning and sterilization.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORNERSTONE SR INSTRUMENTS
Common Device Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC.
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key23507758
Report Number3003120897-2025-00525
Device Sequence Number8973459
Product Code MQP
UDI-Device Identifier00885074464701
UDI-Public(01)00885074464701(11)200715(10)JM19G002
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062073
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number875777
Device Catalogue Number875777
Device Lot NumberJM19G002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2025
Initial Report FDA Received Date11/10/2025
Date Device Manufactured07/15/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-