• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; Implant, endosseous, root-form

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.T.I. BIOTECHNOLOGY INSTITUTE S.L.; Implant, endosseous, root-form Back to Search Results
Model Number IIPECA3785
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/22/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch documentation is checked and no anomalies are detected.Any manufacturing fault is ruled out.After examination of the implant, it can be verified that the implant does not show signs of misuse or signs of malfunction that could have contributed to the event.However, it is found that the drilling protocol used is different from the recommended one.
 
Event or Problem Description
The clinician indicates that he placed the implant on (b)(6) 2024 and loss of osseointegration occurs.Patient with bone quality iii.In the event the patient experienced: bone loss.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
Implant, endosseous, root-form
Manufacturer (Section D)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street,14
minano menor, alava 01510
SP  01510
Manufacturer (Section G)
B.T.I. BIOTECHNOLOGY INSTITUTE S.L.
leonardo da vinci street,14
minano menor, alava 01510
SP   01510
Manufacturer Contact
paula sanchez
leonardo da vinci street,14
minano menor, alava 01510
SP   01510
MDR Report Key23509893
Report Number3004417597-2025-00140
Device Sequence Number18630770
Product Code DZE
UDI-Device Identifier08435389813545
UDI-Public08435389813545
Combination Product (Y/N)N
Initial Reporter CountryEC
PMA/510(K) Number
K211952
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberIIPECA3785
Device Lot NumberBD00252.B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/05/2025
Initial Report FDA Received Date11/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-