• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US APG+ SIZER PIN GUIDE 44+0; EXTREMITY INSTRUMENTS : PIN GUIDES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US APG+ SIZER PIN GUIDE 44+0; EXTREMITY INSTRUMENTS : PIN GUIDES Back to Search Results
Catalog Number 223600002
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that during a left shoulder repair (tsr) surgical procedure it was observed that while trialing the 44mm +0 sizer pin guide from the apg instrument, the surgeon levered on the trial during insertion onto the glenoid and the central peg snapped off.It was removed with no issue.The procedure was completed successfully.There was no delay in the procedure reported.There were no reports of injuries, medical intervention or prolonged hospitalization.No additional information was provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APG+ SIZER PIN GUIDE 44+0
Common Device Name
EXTREMITY INSTRUMENTS : PIN GUIDES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23509982
Report Number1818910-2025-19411
Device Sequence Number18558631
Product Code HWT
UDI-Device Identifier10603295105695
UDI-Public(01)10603295105695
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number223600002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/21/2025
Supplement Date Received by Manufacturer11/11/2025
Initial Report FDA Received Date11/10/2025
Supplement Report FDA Received Date11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-