ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number MA60AC
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that during an intraocular lens (iol) implantation procedure, the lead haptic was improperly folded.The procedure was completed on the same day without any contact or harm to the patient.Additional information has been requested.
 
Additional Manufacturer Narrative
The product was not returned.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The account indicated the use of an unspecified company cartridge.It is unknown if qualified products were used.The company model is only qualified for use in the company cartridges.A handpiece and viscoelastic were indicated which are not qualified for use with this lens, with any of the qualified lens/cartridge combinations.The root cause is most likely related to a failure to follow the (instructions for use) ifu.A handpiece and viscoelastic were indicated which are not qualified for use with this lens, with any of the qualified lens/cartridge combinations.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The lens was returned positioned incorrectly (posterior surface up) in the lens case inside the lens carton.Solution was dried on the lens.One haptic was bent severely at the gusset area at the exit of the haptic insertion area of the optic.The optic was semi-folded and torn from the edge to the optic center.Another area of the optic anterior edge was split.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The account indicated the use of an unspecified monarch cartridge.It is unknown if qualified products were used.The company model is only qualified for use in the company and cartridges.A handpiece and viscoelastic were indicated which are not qualified for use with this lens, with any of the qualified lens/cartridge combinations.One haptic was severely bent/misfolded.This was most likely interpreted as the reported complaint.The root cause is most likely related to a failure to follow the (instructions for use) ifu.A handpiece and viscoelastic were indicated which are not qualified for use with this lens, with any of the qualified lens/cartridge combinations.The ifu instructs: company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.Lenses with a 6.5 mm optic diameter may need to be pre-folded for easy entry into the back of the cartridge.Lens may be pre-folded by gently applying pressure to the edge of the lens while holding the central optic with forceps.The trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF MULTIPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23510245
Report Number1119421-2025-02890
Device Sequence Number9705042
Product Code HQL
UDI-Device Identifier00380652552592
UDI-Public(01)00380652552592(17)260721(11)210723(10)15235901
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/21/2026
Device Model NumberMA60AC
Device Catalogue NumberMA60AC.235
Device Lot Number15235901
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/16/2025
Supplement Date Received by Manufacturer12/01/2025
01/16/2026
Initial Report FDA Received Date11/10/2025
Supplement Report FDA Received Date12/24/2025
02/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV INJECTOR; MONARCH IOL DELIVERY SYSTEM, CARTRIDGE; MONARCH IOL DELIVERY SYSTEM, INJECTOR, UNSPECIFIED; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexMale
-
-