| Catalog Number |
03.033.007 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the trocar and handle are blocked it was not possible to remove the two from each other.The issue was identified pre-operatively.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1 h3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that the weld between the sleeve and the quick coupling was broken.The quick coupling remains attached to the mating device.The fracture surface was examined, and no issues related to material was observed.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces a dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional evaluation was performed several attempts of disassembly were made.The locking mechanism of the hand w/quick-couple was retracted, however, the quick coupling of the protection sleeve ø14 was jammed and could not be disassembled.The complaint condition was replicated; therefore, the allegation can be confirmed.This condition is completely related to the internal damage in locking mechanism of the hand w/quick-couple.The overall complaint was confirmed as the observed condition of the protection sleeve ø14 would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review = > part: 03.033.007-15 lot: 6961p08 manufacturing site: hägendorf release to warehouse: 08-sep-2023.A dhr evaluation was performed for the finished device article # 03.033.007-15 lot # 6961p08, and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|