| Model Number |
4965-15 |
| Medical Device Problem Codes |
Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The lead was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned and analyzed.The distal conductor of the lead developed a fracture due to flexing while in vivo.The distal conductor was fractured at 12.8 cm.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|