| Model Number |
02.12.E0510FL |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Medical device site infection (4845)
|
| Date of Event |
10/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 23 october 2025.Lot 2505013: (b)(4) items manufactured and released on 03-apr-2025.Expiration date: 2030-03-19.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported events during the period of review.Root cause: infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|
Event or Problem Description
|
|
At about 1 month after the primary, the patient came in due to an infection and the pathogen is unknown.The surgeon performed a washout and poly swap.The surgery was completed successfully.
|
| |
|
Search Alerts/Recalls
|