| Model Number |
560BCS1 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the reported display/ power issue was verified during service.Service technician observed that the unit was powered on with no display showing, the fans were also not turning on.And it looks like the system controlled pwa that was replaced last week experienced an issue.The issue was resolved by replacing the system controller pwa.Preventive maintenance was performed as per specification.D9: instrument was serviced at the facility by medtronic field service and was not returned to medtronic service center.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a bio-console 560 controller instrument, it was reported that the user interface display comes on intermittently when the unit is powered on.The instrument was replaced.There was no patient involvement, so no adverse effect occurred.
|
| |
|
Search Alerts/Recalls
|