| Catalog Number |
CS40G |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent 11/10/2025 d4 batch # unk attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: did the device fire? was the device difficult to fire? did the device cut? if the device cut/fired, was the cut line complete? did the device staple? if the device stapled/fired, was the staple line complete? if the device was stapled, was the shape of the staples (b-formation, irregular, unformed)? is the current patient status known? was there any patient consequence or change in the post-operative care of the patient as a result of the event? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the staples that do not penetrate the wall.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 12/11/2025.D4: batch # 873c64.Investigation summary: the product was returned to ethicon endo-surgery for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample revealed that the cs40g device was received with no apparent damage and with a reload present.The reload was received void of staples, with the washer completely cut, drivers exposed and the knife recess below the cartridge deck.In addition, some b-formed staples noted on the reload deck.The device was tested for functionality with a test reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.The event described could not be confirmed as the device performed without any difficulties noted.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Device history review a manufacturing record evaluation was performed for the finished device batch number 873c64, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|