ZIMMER BIOMET, INC. TPRLC XR T1 PPS 15X150MM; PROSTHESIS, HIP Back to Search Results
Catalog Number 51-105150
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2: foreign - event occurred in japan.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that while inspecting circulated items, it was found that the sterile packaging was damaged.There was no patient involvement, and it was reported no further information is available.
 
Event or Problem Description
No further information at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Updated: d9; g3; h2; h3; h6.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility has not been compromised.Review of the provided photos/returned product found the root cause of the reported event can be attributed to transit damage.As no patient or user harm was reported or occurred, the severity level of the reported event is assigned a severity 1 as defined in procedure.A device history review is not required for not reportable, severity 1 events as outlined in the procedure.Complaints are monitored through monthly complaint trending to identify possible adverse trends.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The root cause of the reported event can be attributed to transit damage and a packaging design issue.Upon reassessment of the reported event, it was determined to be not reportable as sterility has not been breached.The initial report should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 15X150MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23511512
Report Number0001825034-2025-03546
Device Sequence Number21265566
Product Code LPH
UDI-Device Identifier00880304517165
UDI-Public(01)00880304517165(17)320404(10)7224205
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K120030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number51-105150
Device Lot Number7224205
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2025
Supplement Date Received by Manufacturer03/10/2026
Initial Report FDA Received Date11/10/2025
Supplement Report FDA Received Date03/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-