Model Number VTICM5 12.1 (-11.00/3.0)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/11/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim# (b)(4).
 
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation.Lens repositioned but problem not resolved.Lens exchanged with different diopter lens.Problem resolved.Cause of the event was not reported.
 
Additional Manufacturer Narrative
H3: device evaluation - lens was returned in microcentrifuge vial dry.Visual inspection found optic deformed and haptic bent.Dimensional inspection found the lens to be within specification.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+VISIAN IMPLANTABLE COLLAMER LENS
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1191 walker ave
monrovia, CA 91016
MDR Report Key23511872
Report Number2023826-2025-01744
Device Sequence Number9666697
Product Code QCB
UDI-Device Identifier00840311305605
UDI-Public(01)00840311305605(11)240728(17)270630
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5 12.1 (-11.00/3.0)
Device Catalogue NumberVTICM5 12.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Supplement Date Received by Manufacturer12/22/2025
Initial Report FDA Received Date11/10/2025
Supplement Report FDA Received Date12/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexFemale
-
-