• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. THUNDERBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. THUNDERBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0520FCS
Medical Device Problem Codes Degraded (1153); Electrical Shorting (2926); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.Based on the results of the investigation, a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the subject device tissue pad was worn.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP TYPE S
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23512110
Report Number9614641-2025-01866
Device Sequence Number21356654
Product Code GEI
UDI-Device Identifier04953170383588
UDI-Public04953170383588
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB-0520FCS
Device Lot Number39K22
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/17/2025
Initial Report FDA Received Date11/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-