As reported, the body of a 5f tempo angiographic catheters was peeling off, appearing as though it was decomposing.There was no reported patient injury.It was further reported that additional devices from the same lots and skus exhibited similar peeling, with approximately nineteen devices affected.It was confirmed that the remaining devices were sterile and exhibited the same failure.Devices are stored in the warehouse along with other medical devices.Ozone is not used to disinfect, and it is not used in any areas such as procedure or storage rooms.Four images were provided, showing blue flakes on a blood-soaked surgical towel, cracks on the catheter body and strain relief, and yellowish discoloration to the hub.The devices will be returned for evaluation.
|
|
As reported, the body of 5f tempo angiographic catheters was peeling off, appearing as though it was decomposing.It was further reported that additional devices from the same lots and skus exhibited similar peeling, with approximately nineteen devices affected.It was confirmed that the remaining devices were sterile and exhibited the same failure.Devices are stored in the warehouse along with other medical devices.Ozone is not used to disinfect, and it is not used in any areas such as procedure or storage rooms.Three sterile devices (¿cath tempo 5f ver 135° 100cm¿) were received for analysis inside their original sealed pouches and placed together in a single plastic bag labeled with complaint numbers (b)(4); one device corresponded to complaint (b)(4), the second device to (b)(4), and the third device to complaint (b)(4), and the product investigation described in this report pertains to complaint (b)(4).The device associated with this complaint was unpacked for evaluation.During visual inspection, the strain relief was found to be partially pulverized and severely cracked, with portions still adhered to the catheter, and yellowish discoloration was observed on the catheter hub; no additional anomalies were detected during the visual examination.Following visual inspection, a microscopic examination was performed on the strain relief, catheter body, and catheter hub areas to further assess the observed conditions, which confirmed severe cracking and partial pulverization of the strain relief, extensive cracking along the length of the catheter body, and confirmation of the partial discoloration observed on the catheter hub, with no additional anomalies identified during microscopic evaluation.Based on the confirmed degradation characteristics observed on the strain relief and the body of the unit, the complaint was escalated for further investigation.The reported ¿catheter (body/shaft) ¿ degradation¿ and ¿strain relief ¿ degradation¿ was observed during the product evaluation.Based on the available information, the specific cause of the degradation could not be determined.Potential contributing factors include storage or handling conditions within the reporting facility or factors associated with the manufacturing process of the unit.Users are trained to inspect products for signs of damage prior to and during use, and any product showing damage is not to be used.The product labeling includes information for safety intended to make users aware of such risks.According to the instructions for use (ifu), ¿store in a cool, dark, dry place.Do not use if package is open or damaged.Do not expose to organic solvents.¿ while this guidance mitigates risks associated with storage and handling, it does not eliminate the potential for degradation.Because a potential link to the manufacturing process could not be excluded, an investigation has been initiated to further evaluate potential contributing factors and implement further actions as appropriate.No additional actions will be taken at this time.
|