| Catalog Number |
24601-B007T |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product or photo was returned by the customer.The customer complaint of leakage could not be verified due to the product not being returned for failure investigation.A device history record review for material# 24601-b007t and lot# 25045188 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 09apr2025.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that bd alaris smartsite texium pump infusion set was leaking the following information was received by the initial reporter with the following verbatim it was reported by the customer that appeared to have been leaking at the connection site of the blue needleless connector and the green texium cap.
|
| |
|
Additional Manufacturer Narrative
|
|
No product or photo was returned by the customer.The customer complaint of leakage could not be verified due to the product not being returned for failure investigation.Although we could not confirm the reported failure, a trend has been identified and actions have been initiated to investigate the issue.Corrective actions taken during this investigation will be implemented in our production process to further increase the robustness and reliability of the device and prevent future occurrences of this type.Bd manufacturing has been informed of this issue.A failure investigation memo and corrective action (trackwise capa pr 11421784) have been initiated.This set was manufactured prior to the implementation date for this capa.A device history record review for material# 24601-b007t and lot# 25045188 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 09apr2025.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
|
| |
|
Search Alerts/Recalls
|