Catalog Number DWT100
Medical Device Problem Codes Unstable (1667); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Joint Laxity (4526); Muscle/Tendon Damage (4532); Insufficient Information (4580)
Date of Event 09/17/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
A revision surgery was performed due to patient dissatisfaction with the outcome of the original surgery.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since medical imaging of the alleged event were provided and confirmed the need for a revision.The device inspection revealed the following: an evaluation of the provided medical imaging by a stryker hcp reveals upward migration of the humerus and is out of alignment with the curvature of the negative glenoid space.This combined with the implant being well fixed and well positioned within the humerus itself is evidence of a rotator cuff deficiency leading to the scheduled revision.A functional and dimensional inspection was not possible since the devices were not returned.However, during manufacturing, functionality test and dimensional inspection were done according to specifications.During the inspection, all devices met the specifications.Based on investigation, the root cause was attributed to a patient factors issue related issue.The event was caused by rotator cuff deficiency.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
A revision surgery was performed due to patient dissatisfaction with the outcome of the original surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM CENTERED HUMERAL HEAD COUPLER TI
Common Device Name
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23514003
Report Number0001649390-2025-00803
Device Sequence Number9667257
Product Code HSD
UDI-Device Identifier00846832085088
UDI-Public(01)00846832085088(11)240529(17)290529(21)AH3237026
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K201315
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberDWT100
Device Lot NumberAH3237
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/13/2025
Supplement Date Received by Manufacturer03/31/2026
Initial Report FDA Received Date11/10/2025
Supplement Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-