• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET DEPTH GAUGE 70MM; KNEE INSTRUMENT: DEPTH GUAGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET DEPTH GAUGE 70MM; KNEE INSTRUMENT: DEPTH GUAGE Back to Search Results
Catalog Number 227460000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the point of the depth gauge broke off while trying to angle the gauge.It happened outside the wound, all pieces were retrieved.No delays.No harm to patient.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (ag0610) provided is not a valid finished goods lot# h11 additional narrative: added: h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET DEPTH GAUGE 70MM
Common Device Name
KNEE INSTRUMENT: DEPTH GUAGE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23514024
Report Number1818910-2025-19506
Device Sequence Number9704702
Product Code HTJ
UDI-Device Identifier10603295109358
UDI-Public(01)10603295109358(10)AG0610
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number227460000
Device Lot NumberAG0610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/22/2025
Supplement Date Received by Manufacturer12/19/2025
Initial Report FDA Received Date11/10/2025
Supplement Report FDA Received Date12/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/15/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-