Model Number VLOCN0604
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a gastric bypass procedure, while suturing, the thread broke in the middle.This issue occurred on ten devices of the same product and lot number.An additional new suture was used without issue.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
D10.Concomitant products: vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy); vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy); vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy); vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy); vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy); vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy); vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy); vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy); vlocn0604, vlocn0604 v-loc* pbt 3-0 blu 15cm v20 (lot a5d1896vy).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that during the gastric bypass procedure, while suturing, the thread broke in the middle.This issue occurred on ten devices of the same product and lot number.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23518491
Report Number1219930-2025-05184
Device Sequence Number18584360
Product Code GAT
UDI-Device Identifier10884521134904
UDI-Public(01)10884521134904(17)280331(10)A5D1896VY(20)04
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K103052
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVLOCN0604
Device Catalogue NumberVLOCN0604
Device Lot NumberA5D1896VY
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2025
Supplement Date Received by Manufacturer02/10/2026
Initial Report FDA Received Date11/10/2025
Supplement Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-