• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION HUMID; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION HUMID; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSXHCP
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Fatigue (1849); Headache (1880); Cough (4457); Sinusitis (4844)
Date of Event 04/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information that the patient alleging coughing, sinus infection, headache and fatigue.There was no report of medical intervention.The manufacturer was made aware of this complaint through a representative of the customer.The device was returned to the manufacturer's product investigation laboratory for investigation.During investigation the manufacturer observed that there were no signs of contamination on the outside of the device.The device was hooked up to a known good power supply and cord.Airflow, heating plate and heated tube were all verified to be operating correctly.The care orchestrator data showed that the patient had a compliance rating, and they used the humidifier.The error log showed that there were no instances of errors on the device.The internal investigation of the humidifier showed that there were no signs of contamination in the humidifier.The internal investigation showed that there were signs of contamination on the inlet port.There were also signs of water ingress in the blower box and on the blower motor and using the fitt tool, the sound abatement foam was depressed and when it was pulled out the foam was stuck, secondary findings water ingress and contamination on inlet port were observed.The manufacturer concludes that unable to address the complaint or symptoms.Manufacturer found evidence of sound abatement foam degradation or breakdown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION HUMID
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key23519539
Report Number2518422-2025-051571
Device Sequence Number13207888
Product Code BZD
UDI-Device Identifier00606959022652
UDI-Public(01)00606959022652
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 11/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDSXHCP
Device Catalogue NumberDSXHCP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2022
Initial Report FDA Received Date11/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-