Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
It was reported that during a laparoscopic-assisted vaginal hysterectomy (lavh) procedure, the ligasure lf5637 retractable l hook experienced a component issue in which the maryland tip would not deploy.They used another handpiece and it worked fine.There was no patient injury.Medtronic's initial evaluation of the incident device found that the shaft of the device is damaged, no pieces are missing.
|
|
Additional information: g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the shaft of the device is damaged, no pieces are missing.The device's jaw was stuck on eschar.Functional testing found that the more frequent cleaning of the jaws could have prevented build up of eschar.The investigation found the shaft was damaged.No pieces detached from the shaft.Analysis of the product showed that the damaged is consistent with an external force on the shaft, very likely user.Had this been an assembly defect, the sample would have been rejected if found prior to shipping.It was reported that the device tip did not deploy.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: inspect the instrument and cords for breaks, cracks, nicks, or other damage before use.Failure to observe this caution may result in injury or electrical shock to the patient or surgical team, or cause damage to the instrument.If damaged, do not use.If the instrument shaft is visibly bent, discard and replace the instrument.Keep the instrument jaws and l-hook clean.Build-up of eschar may reduce the sealing, cutting, dissection effectiveness, and may heat jaws to the point of damaging jaw insulation.Wipe jaw surfaces and edges with a wet gauze pad as needed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|